Warranty: | 1 Year |
---|---|
Structure: | Portable |
Material: | Plastic |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Latest Microprocessor Controlling System
One button auto zero calibration and 100%T calibration.
128*64 LCD display, direct readout of Transmittance, Absorbance and Concentration.
Precision 1200 lines/mm Grating Monochromator
Ensure high photometric resolution, high accuracy and low stray light.
Easy Data Output by USB/Parallel Port
USB Port: Connect to computer and operate through PC software. ( PC software is optional.)
Parallel Port: Connect to micro thermal printer to print test data. ( Micro thermal printer is optional.)
Wide Sample Chamber Enables Multi-functions
120mm wide sample chamber and long optical path monochromator provide enough space for optional accessories.
Four Display Modes; One-Button Quick Switching
Transmittance, Absorbance, Concentration and Factor (Coefficient) mode.
Model | V-5000 |
Optical System | Single Beam, grating 1200 lines/mm |
Wavelength Range | 325-1000nm |
Spectral Bandwidth | 4nm |
Wavelength Accuracy | ±2nm |
Wavelength Repeatability | ≤0.5nm |
Wavelength Setting | Manually |
Photometric Accuracy | ±0.5%T |
Photometric Repeatability | ≤0.2%T |
Photometric Range | -0.3-3A, 0-200%T, 0-9999C |
Photometric Mode | T, A, C, F |
Stray Light | ≤0.2%T at 360nm |
Display | 128*64 Dots LCD |
Detector | Silicon Photodiode |
Output | USB Port & Parallel Port |
Light Source | Tungsten Lamp |
Power Requirements | AC 85~250V |