This instrument can be used in turbidity monitoring ofsecondary water supply, waterworks, municipal pipe network,and can also be used in industrial process water qualitymonitoring, circulating cooling water, turbidity monitoring ofwater filtered from the membrane and other environmentsthat require high accuracy of turbidity detection.
Features:
-Drinking water quality turbidity detector,the detection limit is as low as 0.015NTU,has high measurement accuracy and resolution.
-Exclusive flow cell with built-in steady flow and defoaming system to reduce theinterference of flow rate and air bubbles on the measured value.
-Using laser light source, long life cycle, maintenance-free light source.
-Have anti-condensation,anti-dewing function.
-Built-in screen to display the measured value, uses can check the data quickly.
Measuring range(NTU) | 0-20 other ranges are optional |
Resolution(NTU) | 0.001,0.01 |
Indication error(NTU) | ±2% or ±0.02 NTU whichever is greater |
Zero Drift | ≤0.015NTU |
Calibration | Water sample calibration; Standard solution calibration |
Sample Flow Rate | 300mL/min≤X≤800mL/min |
Operating temperature | 0-50ºC(no freezing) |
Storage temperature | -10~60ºC |
Power supply | 12V/0.03A |
Digital interface | MODBUS/RS485 |
IP grade | IP67 |
Main material | ABS |
Installation method | Flow cell installation |
Dimension | 367*213*115mm |
Cable length | 1.5m(default) |